Model | X Travel (mm) | Y Travel (mm) | Z Travel (mm) | MPEe (μm) | MPEp (μm) |
---|---|---|---|---|---|
Libra686 | 600 | 800 | 600 | 0.6+L/500 | 0.8 |
Libra8106 | 800 | 1000 | 600 | 0.9+L/500 | 1.0 |
Libra10128 | 1000 | 1200 | 800 | 1.0+L/500 | 1.5 |
Libra12167 | 1200 | 1600 | 700 | 1.1+L/500 | 1.5 |
Libra121510 | 1200 | 1500 | 1000 | 1.2+L/500 | 1.5 |
Model: Libra686
X Travel: 600 mm
Y Travel: 800 mm
Z Travel: 600 mm
MPEe: 0.6+L/500
MPEp: 0.8
Model: Libra8106
X Travel: 800 mm
Y Travel: 1000 mm
Z Travel: 600 mm
MPEe: 0.9+L/500
MPEp: 1.0
Model: Libra10128
X Travel: 1000 mm
Y Travel: 1200 mm
Z Travel: 800 mm
MPEe: 1.0+L/500
MPEp: 1.5
Model: Libra12167
X Travel: 1200 mm
Y Travel: 1600 mm
Z Travel: 700 mm
MPEe: 1.1+L/500
MPEp: 1.5
Model: Libra121510
X Travel: 1200 mm
Y Travel: 1500 mm
Z Travel: 1000 mm
MPEe: 1.2+L/500
MPEp: 1.5
Fixed Bridge Ultra-Precision CMM —— Libra
The Fixed Bridge Coordinate Measuring Machine is designed for ultra-precision measurement and supports touch-trigger, analog scanning, and laser scanning modes. Its fixed bridge structure delivers exceptional rigidity and stability, making it ideal for high-performance measurement across various part sizes and complexities.
Key Features:
The fixed bridge design offers exceptional stiffness. Combined with imported granite materials, it minimizes deformation caused by mechanical stress or temperature changes—ensuring long-term accuracy, stability, and performance.
A fully enclosed air-bearing layout provides excellent resistance to environmental interference and significantly reduces Abbe errors, maintaining precision even at high scanning speeds.
Equipped with ultra-high-resolution floating glass scales, the machine ensures dimensional accuracy and repeatability while minimizing thermal expansion effects on measurement stability.
The moving platform uses a high-precision flexible ball screw drive system, delivering outstanding motion stability and accurate directional positioning.
Optional precision vibration isolators are available based on machine size and payload requirements to ensure optimal measurement stability and accuracy.
Environmental Conditions
Operating Temperature: 20℃ ±1℃
Temperature Gradient: ≤1℃/h, ≤2℃/24h, ≤1℃/m
Humidity: 55% – 65% RH
Power Supply: AC 220V ±10%, 15A
Air Pressure: 0.6 – 0.8 MPa
Air Consumption: >230 NL/min
Maximum Load Capacity: 600 kg
Vibration Tolerance:
At frequencies below 10 Hz: amplitude must be ≤2 μm P–P
At 10–50 Hz: acceleration must be ≤0.4 gal